Bovie introduces the Aaron OR PRO 300 Electrosurgical Generator. The system was designed in conjunction with OR clinicians to provide a safe, user-friendly generator to meet all operating room procedure performance demands. This innovative energy-based device offers 13 modes of surgical energy delivery. Special modalities include the advanced Bovie Bipolar, Auto Bipolar options and special Laparoscopic modes. These modes were included to assist surgeons in tailoring all electrosurgery needs for their variety of surgical applications and user preferences.
What's Included:
- Aaron OR|PRO 300 Electrosurgical Generator
- 110 VAC Hospital-Grade Power Cord (15ft.)
- 220 VAC Hospital-Grade Power Cord (15ft.)
- User's Guide in CD Format (MC-55-230-001)
Monopolar | Watts | Vpeak Max Voltage | Crest Factor (Typical) | Frequency (KHz) | Cut I | 300 | 1000V | 1.7 +/- 20% | 490 kHz +/- 4.9 kHz | Cut II | 300 | 750V | 1.7 +/- 20% | 490 kHz +/- 4.9 kHz | Blend (1) | 200 | 1320V | 1.8 +/- 20% | 490 kHz +/- 4.9 kHz | Blend (2) | 200 | 1475V | 2.0 +/- 20% | 490 kHz +/- 4.9 kHz | Blend (3) | 200 | 1650V | 2.2 +/- 20% | 490 kHz +/- 4.9 kHz | Blend (4) | 200 | 1870V | 2.4 +/- 20% | 490 kHz +/- 4.9 kHz | Pinpoint Coagulation | 120 | 1800V | 3.1 +/- 20% | 490 kHz +/- 4.9 kHz | Spray Coagulation | 120 | 4000V | 6.0 +/- 20% | 350 to 450 kHz | Gentle Coagulation | 120 | 450V | 1.6 +/- 20% | 490 kHz +/- 4.9 kHz | Bipolar | Watts | Vpeak Max Voltage | Crest Factor (Typical) | Frequency (KHz) | Macro Bipolar | 80 | 600V | 1.5 +/- 20% | 490 kHz +/- 4.9 kHz | Micro Bipolar | 80 | 450V | 1.5 +/- 20% | 490 kHz +/- 4.9 kHz | Standard Bipolar | 50 | 300V | 1.5 +/- 20% | 490 kHz +/- 4.9 kHz | Bovie Bipolar | 225 | 250V | 1.5 +/- 20% | 490 kHz +/- 4.9 kHz | Auto Bipolar | 80 | | | 490 kHz +/- 4.9 kHz |
| Contact Monitoring Systems | BovieNEM | Contact Monitoring Systems Range/Trip Point | 10 ohms or less over entire operating range (10C to 40C) | Cooling System | Passive Cooling | Technology | Isolated RF Output |
| Audible/Visible Output Meter | Yes | Pre-set Feature | Yes | Current Meter | Yes | Monopolar HP Ports | 2 | Dual Activation Tone Alert | Yes | Connector Ports/Power Switch | Black |
| Weight: | <9.07 kg (<20 lbs) | Dimensions: | W - 37.5 cm (14.75 in) H - 16.5 cm (6.5 in) D - 46 cm (18.1 in) | Warranty | 4 years, Additional Warranty Available
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Features
Bovie Bipolar (BBp)
Bovie Bipolar is a pulsing coagulation for quick and powerful coagulation of tissue. The pulsing performance can be tailored specifically to achieve the optimal surgical effect. The feature is compatible with Bovie's Micro, Macro & Standard Bipolar modes and can be set to surgical preferences for critical procedures. Auto Bipolar Intuitive Technology
Auto Bipolar offers immediate and simplified tissue coagulation with minimal tissue damage. This feature automatically activates, coagulates and deactivates for precise tissue coagulation and eliminates the need for a footswitch. This feature is compatible with Bovie's Micro, Macro and Standard Bipolar modes and can be set to surgical preferences for critical procedures.
Surgical Memory Technology The Aaron OR|PRO 300 recalls the last used power and mode. The unit's user preset can recall up to 10 presets and up to 6 bipolar output preferences. Adjustable Power during Activation During all activations, the power can be adjusted up or down in variable steps, giving surgeons more control at the surgical site. Patient Safety Features
-
BovieNEM - This return electrode contact quality monitoring system is designed to minimize patient burns at the return electrode site.
-
Bovie FDFS - The tissue sensing technology measure tissue 5 million times a second and adjusts to varying impedances 5000 times a second for accurate energy delivery.
-
A visual Ammeter indicates real time current (mA) flowing through the patient.
Energy Settings
CUT
Cut I - A general cutting with a standard RF waveform used in open and endoscopic surgical procedure, Cut I is for use in a wide range of tissue with minimal hemostasis.
Cut II - Tailored specifically for Laparoscopic procedures, Cut II offers a precisely managed cutting energy that limits output for the minimizing of the potential risk of capacitive coupling and alternate site burns due to weakened insulation.
Blend (Level 1, 2, 3 & 4) - Blended cutting offers varying levels of hemostasis during a cut providing increased control of bleeding at the surgical application site.
Blend Level
|
Duty Cycle
|
1
|
75%
|
2
|
62.5%
|
3
|
50%
|
4
|
37.5%
|
COAGULATION
Pinpoint - A general coagulation mode delivering rapid desiccation effects with minimal tissue damage.
Spray - Recommended for hemostasis in highly vascular or large surface areas and for non-contact tissue coagulation to control superficial bleeding. Spray mode permits dual, simultaneous activation of two monopolar accessories with audible dual activation tone alert.
Gentle - Gentle mode offers more concentrated coagulation than in other modes. For instance, when coagulation is necessary in short amounts of time, reduced electrode carbonization is provided. BIPOLAR
Standard - Bipolar coagulation which gently coagulates tissue for minimal tissue damage reducing carbonization and sticking.
Macro - Intense bipolar output allowing for aggressive coagulation, with cut-like effects over a wide range of tissue.
Micro - Fine bipolar coagulation for use in extremely gentle and precise surgical applications.
Bovie Bipolar - Exclusive to the Aaron OR|PRO 300 and the IDS-310, Bovie Bipolar is a pulsing bipolar coagulation for quick and powerful coagulation of tissue. Pulsing performance can be tailored specifically to achieve the optimal surgical effect.
|
|
|
|
|
|
|